ENGINEERED FOR OPTIMAL PERFORMANCE AND FLEXIBILITY
MicroScan Systems provide a true growth-based ID/AST system for microorganism identification and susceptibility testing. The WalkAway plus System includes LabPro Information Management software designed to simplify workflow and minimize technologist interaction while accommodating differing regional and institutional environments through extensive customization features. The customizable LabPro AlertEX software automates detection of atypical results for quick reporting and directs staff to the most appropriate action based on customized institutional procedures.
MicroScan WalkAway Product Features:
ONE SYSTEM FOR AUTOMATION OF RAPID AND CONVENTIONAL TESTS
Choose conventional overnight ID/AST, and/or specialty and rapid ID testing simultaneously on one instrument without sacrificing accuracy
Automatic incubation, test interpretation, and reagent control frees staff for other tasks
Available in both 40- and 96-panel capacity models
ENHANCED FEATURES COME STANDARD
Delivers gold-standard accuracy for microorganism identification and susceptibility testing.
Enables simultaneous processing of conventional, rapid, and specialty panels on a single, automated platform.
Delivers accurate emerging resistance detection for the toughest pathogens, including VISA, VRSA, and MRSA.
Processes Rapid and Specialty ID panels for reduced turnaround when speed matters.
Beckman Coulter works closely with you to help your microbiology lab reach new levels of operational efficiency.
Specifications
Physical
WalkAway-40 plus: 29" H x 38.5" W x 34" D; 375 lbs
WalkAway-96 plus: 37" H x 38.5" W x 34" D; 400 lbs
Color: Two-tone gray
Interface connector:
Serial interface connector (two): daisy-chained RS 232
"In" port toward computer, "out" port away from computer
Power Requirements
Line voltage: 100/120/220/230/240 VAC selected by internal jumpers
Line frequency: 50/60 Hz ±5%
Line current:
9 A (nominal), 11 A (maximum) for 100/120 VAC operation
4.5 A (nominal), 5.5 A (maximum) for 220/230/240 VAC operation
Power loss protection: Circuitry provided to ensure fail-safe operation; power interruption will not result in a hazard; automatic restart
Heat generation: 3,400 BTU/hour
Recommended Laboratory Environmental Conditions
Temperature: (operating) 15°C to 27°C
Humidity: 30% to 80% RH (no condensation)
Safety Requirements
Ground impedance: 0.1 Ω (maximum)
Dielectric withstand voltage: No dielectric breakdown (at 1,400 VDC for one minute)
Leakage current: <0.5 mA
Removable power cord:
3-conductor, 14 AWG (100/120 V)
3-conductor, 16 AWG (220/230/240 V)
Interlocks: Provided on panel access door, reagent dispense drawer and service hatch
Power plug: NEMA type 5-15P/IEC 60320 type C14
Standards: Certified to conform to UL STD 61010-1 (second edition), CAN/CSA STD C22.2 Nos. 61010-1 (2004), 61010-2-081 (2004), 61010-2-101 (2004), ISO 9000, CE, IVDD, EMC STDs IEC61326-1:2005-12, and IEC61326-2-6:2005-12
Overcurrent Protection
Fusing: Primary fuses F1 and F2, IEC (5x20 mm size) time lag/slowblow type, 250 V rated, 12 A (100/120 V)/ 6.3 A (220/230/240 V)
Panel capacity for WalkAway-40 plus System: 40 MicroScan panels with 96 wells each of any type compatible with the incubator environment; up to eight panel towers with five panels each
Panel capacity for WalkAway-96 plus System: 96 MicroScan panels with 96 wells each of any type compatible with the incubator environment; up to eight panel towers with 12 panels each
Incubation temperature stability: ±1°C
Reagent type capacity: 10 reagents total
Reagent dispense volume (μl): 15 to 75, depending on reagent dispensed
Oil dispense volume (μl): 60 to 150
Operator interface: Front-panel controls and PC keyboard, monitor and LabPro software
Be the first one to review this product.
Your review is submitted for approval.
No downloadable files available
Subscribe to Our Newsletter
Subscribe to our newsletter & get notification about discounts.
Subscribe to Our Newsletter
Subscribe to our newsletter & get notification about discounts.
By subscribing to our newsletter you agree to our Privacy Policy.
The website uses cookies to ensure you get the best experience on our website.